Detector Correction

 

 

While using fluorescence data from a detector array, some detectors can be noisier than the others, and thus need to be excluded from the averaging.  It is always advisable to check the Fluorescence from each of the individual crystals. Check the box on the left of "Plot F" and then choose the first fluorescence detector (“ge1” for this example) from the drop-down menu as shown.

 

 

Place Pointer over the marked region to zoom in

Place Pointer over the marked region to zoom in

Place Pointer over the marked region to zoom in

 

 

Observe the plot after clicking OK and if it is noisy, the readings of that particular crystal can be ignored. To rapidly check the spectra from the other detector crystals, click repeatedly on the "down" button next to the drop-down menu on the LX Panel. (For an alternative method, see note below.) In the example used, data from ge3, ge4, ge8 and ge12 were deemed noisy and omitted from averaging after observing their plots. To do this, click on "Detector Info Table" on the left of the LX Panel. In the table that appears, set the "det_wt" of ge3, ge4, ge8 and ge12 to 0.

 

 

Note: If you prefer to view or print a plot containing overlayed spectra from all the detectors (rather than viewing them one at a time), from the Main Menu go to EXAFS123> LX Additional Routine > LX Special Graph routine. From the Dialogue BIx that appears, select "Fluorescence by Scan" and hit "Continue".

To see the effects of the changes you made, click "Update Graph after changing Det. or File" button on the center of the LX Panel.