Calibration
The energy scale might need to be corrected based on the
calibration data (if in-line calibration data was obtained). On the LX Panel, click on "Calibration
Graph". From the drop down menu in the dialogue box that appears, select
the option corresponding to the reference sample used. As KMnO4
was used in the given experiment, "KMnO4 powder with peak
at 6543.3 eV" is selected.
After hitting Continue, the calibration graph is seen and
on the lower left part of this window, you see the value of "Ecorr" which, for this experiment, is 1.21 eV. To "correct" the plot using this, simply check
the box next to "Ecorr" on the LX Panel.